Orta boy slim-line otoklavlar

Orta boy slim-line otoklavlar

GS Hospital AutoclavesFor CSSDs, OR & Medical Centers

160 liter (1 StU) or 250 liter (2 StU) chambers
Pass-through between sterile & non-sterile areas
Single and Double Door
Sterilization of wrapped and unwrapped equipment
Cost-effective pricing
Hospital grade performance

GS-LINE HOSPITAL AUTOCLAVES

GS autoclaves are designed for Hospital CSSD, OR, and Medical Centers with limited budgets that need reliable sterile processings. We have maintained Tuttnauer’s high quality for hospital grade autoclaves, providing reliable, heavy-duty high performance machines. The rounded chamber and coiled jacket reduce the manufacturing costs which allow for more affordable autoclaves.

MORE AFFORDABLE

For Facilities with Limited Budget

• Rounded chamber and coiled jacket reduces the manufacturing cost of the autoclave
• Tuttnauer passes that benefit on to our customers

NARROW DESIGN

Designed for Limited Space

• Narrow 75cm wide autoclaves for more autoclaves in a limited area
• More autoclaves reduce downtime risks by ensuring continued sterile  processing

SIMPLIFIED DESIGN = LOW MAINTENANCE

Hospital Grade Performance with Simplified Design

• Less parts and reduced complexity
• Light weight and simplified installation with flexible connections
• Low long-term maintenance costs

CONVENIENT 1 WALL PASS-THROUGH INSTALLATION

• Wall Installation between non-sterile and sterile rooms
• Includes side panels for 1-wall installation
• Additional service area not required
• Can also be installed in a 2 wall separation

FEATURES

  • Automatic locking hinged doors
  • 7” Touch screen on both sides
  • 18kW built-in steam generator
  • Built-in wheels
  • Built-in water-ring vacuum pump
  • 316L Stainless steel chamber
  • 304L Stainless steel housing
  • Built-in printer

5075GS AUTOCLAVE

• 160 Liter chamber   • 1 StU basket   • 2 ISO baskets

5075GS AUTOCLAVE

• 160 Liter chamber   • 1 StU basket   • 2 ISO baskets

TECHNICAL SPECIFICATIONS

Model Chamber Dim.
ø x L (mm)
Chamber Vol. (Liters) Tray Dimensions
WxHxD (mm)
Baskets Power (kW) Voltage/Freq.
Volts/Hz
50125GS-1A (1 door)
50125GS-2A (2 doors)
500 x 1250 250 2 x [310 x 21 x 610]

2 x [471 x 21 x 610]

2 StU

4 ISO

18 kW 3 ph, 208/230/400 Volts
1ph, 230 Volts* 50/60 Hz
5075GS-1A (1 door)
5075GS-2A (2 doors)
500 x 810 160 310 x 21 x 773
471 x 21 x  773
1 StU
2 ISO
18 kW 3 ph, 208/230/400 Volts
1ph, 230 Volts* 50/60 Hz

* requires switch box – optional
Note: The 50125GS requires a compressed air supply of 6 Bar to operate pneumatic valves.

SAFETY

Safety for personnel, autoclave and load are priority in the design, construction and operation of any Tuttnauer autoclave. Tuttnauer is committed to the highest industry safety standards and directives to ensure safety not only for your employees operating the autoclaves but also for your laboratory and the loads being sterilized.

  • A safety device prevents the operator from opening the door when the chamber is pressurized
  • Steam will not enter the chamber when the door is open
  • A cycle cannot start if the door is open or not properly locked
  • The door cannot unlock until liquid temperature reaches the predetermined end temperature
  • Two independent flexible PT100 temperature sensors to prevent over-boiling of liquids and explosions of bottles